PSC Magellan 8500 Specifikace Strana 111

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 480
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 110
Product Reference Guide 5-9
Shift Test
The Shift Test checks to ensure that items placed anywhere on the weigh-
ing surface of the scanner/scale ar
e weighed properly. Refer to Figure 5-2
when performing this test.
1. Place and remove in succession, a ten-pound load on the center of
each of the
four quadrants (1, 2, 3, and 4 in Figure 5-2) of the
scanner/scales weigh platter. Verify that a reading of between 9.99
and 10.01 lb registers under load and that the display returns to
0.00 between each load.
2. After verifying the accuracy of each quadrant of the weighing sur-
face, remove all weight from the sc
ale. You have completed the
Shift Test.
Figure 5-2. Shift Test
12
34
1
2
4
3
(Models vary)
Zobrazit stránku 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 479 480

Komentáře k této Příručce

Žádné komentáře